Manufacturer by Type

Products by Category



NF ZM2371 LCR Meter 1mHz to 100kHz

PDFPrintEmail
NF ZM2371 LCR Meter 1mHz to 100kHz
NF zm2371 LCR MeterNF zm2371
Price:
Manufacturer: NFCorporation
Description

NF ZM2371 LCR Meter

A LCR meter measures impedance (inductance, capacitance and resistance) at spot frequencies. It is used for testing components and materials in R&D applications and in component manufacturing.

  • Basic accuracy 0.08%
  • Measurement frequency 1mHz to 100kHz
  • Measurement speed Max. 2ms 

ZM2371/ZM2372 (Photo shows the ZM2371) 

Primary parameters 
Displays measured values for Lp, Ls, Cp, Cs, Rp, Rs, |Z|, |Y|, G.
Automatic selection is possible. 

* Secondary parameters 
Displays measured values for Q, D, theta, X, B, Rs, Rp, G, Lp, Rdc (DC resistance).
Automatic selection is possible. 
* Measurement frequency 
Select one of the four frequencies (120 Hz, 1 kHz, 10 kHz, and 100 kHz) or input a numeric value (5-digit resolution).
LCR meter ZM2353

Equivalent circuits(Primary parameter) 
For the L, C, and R, specify the equivalent circuit. Automatic selection is possible. 

* Measurement range 
Automatic selection (AUTO) and manual selection (HOLD) can be switched.The default setting is automatic selection.
* Measurement speed
Switchable between 5 levels from 2 ms to 501 ms.
mark Basic accuracy: 0.08 %, display resolution of 6 digits (max.)

 

Measurement speed: Max. 2 ms at 1 kHz
space
Measurement speed is selectable between 5 levels: RAP (rapid) , FAST, MED, SLOW and VSLO (very slow). When set to RAP, high-speed measurement at 2 ms (1kHz) or 10ms (120 Hz) can be performed.

 

Measurement frequency: 1 mHz to 100 kHz, 5-digit resolution

 

Measurement signal level: Max. 5 Vrms, 3-digit resolution

 

Measurements parameters: Lp, Ls, Cp, Cs, Rp, Rs, |Z|, |Y|, G, Q, D, theta, X, B, Rdc

 

mark Internal DC bias: 0 V to +2.50 V

 

mark Constant-voltage/constant-current drive (ALC:Auto Level Control)

 

mark DC resistance measurement

 

4-terminal contact check function and a handler interface for production lines (ZM2372)
space
ZM2372 is equipped with contact check function to prevent mis-measurement or incorrect sorting due to defective contact between the measurement tip and component. It checks contact with 4 measurement terminals, and if a contact defect occurs, it halts measurement/sorting and determines the contact defect. This function contributes to improved yield.
space
contact check

Step1:
Checks whether there is continuity along the dotted lines (contact check).
Determines there is a defect if contact is defective.

ya
Once the contact failure is detected, ZM2372 allows the measured value “NC” to appear without performing normal measurement. 
The additional time necessary for 4-terminal contact check is approx. 4 ms.
ya
Step2: Measures along the dotted lines (main measurement)
 

 
 

 

mark Other functions: 
Correction functions (Open, Short, Load and Cable length), Deviation measurement, Comparator function etc.

 

mark Interfaces
space
  • RS-232
  • USB
  • GPIB 
    (Standard for ZM2372, not provided for ZM2371)
  • Handler interface
    (Standard for ZM2372, not provided for ZM2371) 

zm2372_rear panel

 

mark Application software